Image Not Available

Overview

The 27333 allows customers to send their EC155, CPEC200, CPEC300, CPEC306, or CPEC310 to Campbell Scientific and have Campbell Scientific perform a zero and span; an RMA (Return Material Authorisation) is required. The zero and span should be done when ambient conditions change significantly (i.e., seasonally).


Read More

Detailed Description

With time, the EC155, CPEC200, and CPEC3XX gas analyzers can experience some drift in signal offset and gain, especially under varying environmental conditions. Thus, a zero and span procedure (i.e., a simple two-point calibration) should periodically be performed on the analyzer.

The zero and span is usually performed in the order of several weeks or months. Anytime there is a significant change in ambient conditions (e.g., change in seasons) or when the instrument appears to have drifted, a zero and span should be done.


Compatibility

Please note: The following shows notable compatibility information. It is not a comprehensive list of all compatible products.

Miscellaneous

Product Compatible Note
CPEC200 (retired)
CPEC300-Series (retired)
CPEC306
CPEC310
EC155


Privacy Policy Update

We've updated our privacy policy.  Learn More