The HMP155A is an air temperature and relative humidity (RH) sensor that provides reliable measurements for a wide range of applications. It can be used as part of a weather station system or as a single instrument. It replaces the HMP45C, which is still available but will be retired in late 2010. It uses the 41005-5 radiation shield, which is different from the one used by the HMP45C.
Specifications
Specifications
- Electromagnetic Compatibility: Complies with EMC standard EN61326-1 Electromagnetic
- Filter: Sintered PTFE
- Housing
- Material: PC
- Classification: IP66
- Voltage Output: 0 to 1 Vdc
- Average Current Consumption: ≤3 mA (analog output mode)
- Operating Voltage: 7 to 28 VDC
- Settling Time at Power Up: 2 seconds
Relative Humidity (RH)
- Sensor: HUMICAP®180R
- Measurement Range: 0.8 to 100% RH, non-condensing
- Response Time*: 20 s (63% RH), 60 s (90% RH)
- Factory Calibration Uncertainty (+20°C)**
- 0 to 40% RH: ±0.6% RH
- 40 to 97% RH: ±1.0% RH
- Accuracy (including non-linearity, hysteresis & repeatability)
- +15° to +25°C: ±1% RH (0 to 90% RH) ±1.7% RH (90 to 100% RH)
- -60° to -40°C: ± (1.4 + 0.032 × reading) % RH
- -40° to -20°C: ± (1.2 + 0.012 × reading) % RH
- -20° to +40°C: ± (1.0 + 0.008 × reading) % RH
- +40° to +60°C: ± (1.2 + 0.012 × reading) % RH
Air Temperature
- Temperature Sensor: Pt 100 RTD 1/3 class B IEC 751
- Measurement Range: -80° to +60°C
- Accuracy with Voltage Output
- -80° to +20°C: ±(0.226 - 0.0028 x temperature)°C
- -80° to +60°C: ±(0.055 - 0.0057 x temperature)°C
- Entire Temperature Range: see graph in product brochure
*The response time for the RH specification is for the HUMICAP®180R© at 20°C in still air with sintered PTFE filter.
**The factory calibration uncertainty is defined as ±2 standard deviation limits. Small variations possible; see also calibration certificate.