If a CR1000 will not pass self-calibration, what can be done to fix that?

A voltage outside of ±5 V present on one of the analog channels can cause self-calibration to fail. Disconnect all sensors, cycle power, and check if the error remains. If self-calibration fails with no sensors attached, consult an application engineer about sending the data logger to Campbell Scientific for repair.